Abstract
In this study, McGregor’s analytical method was applied to calculate the absolute efficiency of the TCS-172 GSSM equipped with a 1 in. ´ 1 in. NaI(Tl) detector. Calculations were performed for 662 keV and 1250 keV gamma rays emitted from Cs-137 and Co-60 point sources, respectively. The calculated efficiencies were compared with experimental measurements conducted at different source-to-detector distances. The calculated data were fitted over 0.5–5.0 cm and evaluated at the experimental distances. The differences ranged from −7% to 6% for 662 keV and −9% to 6% for 1250 keV, with a maximum absolute difference of 9%. The results showed that the absolute efficiency decreases with increasing distance and is higher at 662 keV than at 1250 keV due to the greater interaction probability of lower-energy gamma rays in NaI(Tl). Good agreement was observed between the calculated and measured efficiencies, confirming the applicability of McGregor’s analytical method to the TCS-172 detector. The proposed approach provides a simple and rapid tool for estimating the absolute efficiency of portable NaI(Tl)-based gamma survey meters.